Testing & COAs

Mass Spectrometry for Peptide Identity: ESI, MALDI and What the Peaks Mean

HPLC says how pure a sample is; mass spectrometry says what it is. A guide to ESI and MALDI spectra, charge states, deconvolution and mass tolerance on a peptide COA.

Scientist adjusting an ion-trap mass spectrometer in an analytical laboratory
Photo: Mwillett4000, Wikimedia Commons · Public domain

Purity and identity are different questions. A sample can be 99% pure and still be the wrong peptide. Mass spectrometry (MS) is how a laboratory confirms that the molecule in the vial has the molecular weight its sequence predicts.

The principle in one paragraph

A mass spectrometer turns molecules into gas-phase ions, separates them by their mass-to-charge ratio (m/z) and counts them. Peptides are usually ionised by adding protons, so a peptide of mass M carrying n extra protons appears at roughly (M + n) ÷ n. The instrument reports a spectrum: m/z on the x-axis, signal intensity on the y-axis.

ESI: many peaks, one molecule

Electrospray ionisation (ESI) sprays the dissolved sample through a charged needle. Larger peptides pick up several protons, so one molecule appears as a series of peaks: [M+2H]²⁺, [M+3H]³⁺, [M+4H]⁴⁺. Software deconvolutes that envelope back to a single neutral mass.

ESI is often coupled directly to HPLC (LC-MS). That gives a mass for the main chromatographic peak specifically, and often for the impurity peaks next to it, which is very useful for identifying deletion sequences or oxidation products.

MALDI-TOF: usually one peak

Matrix-assisted laser desorption/ionisation (MALDI) co-crystallises the sample with a matrix and fires a laser at it. Peptides mostly pick up a single proton, so the spectrum shows a dominant [M+H]⁺ peak, sometimes with sodium [M+Na]⁺ and potassium [M+K]⁺ adducts 22 and 38 Da higher. A time-of-flight (TOF) analyser measures m/z from how long ions take to reach the detector.

What to check on the certificate

Item What good looks like
Theoretical mass Calculated from the sequence, stated as monoisotopic or average
Observed mass Within the instrument’s stated tolerance of the theoretical value
Spectrum image Present, with the main peak labelled
Method ESI or MALDI, instrument type, mode

Be clear on monoisotopic vs average mass. The monoisotopic mass uses the most abundant isotope of each element; the average mass weights all isotopes. For a 4–5 kDa peptide they differ by a few daltons, so a report must compare like with like.

Common modifications you can spot by mass

  • +16 Da: oxidation, typically methionine or tryptophan
  • +1 Da: deamidation of asparagine or glutamine (subtle; needs good resolution)
  • −17 or −18 Da: pyroglutamate formation from N-terminal Gln or Glu
  • +114 Da (TFA), +22 Da (Na): adducts, not impurities in the vial

Identity plus purity

MS confirms what the molecule is; HPLC quantifies how much of the sample it is. A complete COA has both. For the purity side, read HPLC purity explained; for the certificate as a whole, see how to read a peptide COA.

For in-vitro laboratory research only. Not for human or veterinary use. This article summarises laboratory practice and published research. Nothing here is advice about use in people or animals.

Related articles

All articles